Membrane Filter 47 mm
Membrane Filter 47 mm is a porous filter disc used to retain and count microorganisms in water and other liquid samples by the membrane filtration method. After the sample is filtered, the filter is placed on a selective agar and the colonies growing on its surface are counted.
Special pricing in line with technical specifications is offered for institutional and hospital purchases.
Technical specifications
About the product
Membrane Filter 47 mm is a thin, porous filter disc with a diameter of 47 mm used to retain and count microorganisms in water and other liquid samples by the membrane filtration method. The sample is drawn through the filter and bacteria remain on its surface; the filter is then placed on a selective agar medium and the colonies that grow on the surface are counted after incubation. Pipette and filter options are listed under Pipettes and Membrane Filters.
Key features
- Diameter: 47 mm; a widely used size in membrane filtration units
- Pore size: Specified in the quote; in water microbiology a 0.45 µm pore size is commonly used for bacterial counts
- Gridded surface: On common models, printed grid lines make colony counting easier
- Handling: Transferred directly onto the agar surface after filtration
- Material, sterility and packaging: Specified in the quote
Applications
- Counting indicator organisms in drinking water, process water, pool water and bottled water
- Coliform bacteria and E. coli: the filter is placed on Chromogenic Coliform Agar (CCA) (ISO 9308-1)
- Intestinal enterococci: the filter is incubated on Slanetz Bartley Agar (ISO 7899-2)
- Liquid samples with a low microbial load that must be filtered in large volumes
How to use
- Sterilise or disinfect the filtration unit (funnel, support and vacuum source) as required by the method.
- Hold the filter with flat-tipped, previously disinfected forceps and place it on the support with the gridded side up; do not touch the filter by hand.
- Fit the funnel, pour in the sample volume stated by the method and filter under vacuum; rinse the funnel wall with sterile diluent if needed.
- Release the vacuum, remove the funnel, lift the filter with the forceps and lay it on the agar starting from one edge so that no air bubbles remain underneath.
- Incubate the plate at the temperature and for the time stated by the method; count the characteristic colonies using the grid squares and calculate the result per filtered volume.
- Include a negative control filter with sterile water in each series.
What to consider when choosing
- Pore size: Choose the pore size required by the standard or method you apply.
- Diameter fit: Check that the 47 mm filter fits your filtration unit and the Petri dishes you use.
- Grid and colour: A grid simplifies counting; a filter colour that contrasts with the colonies speeds up reading.
- Sterility and packaging: Sterile, individually packed filters make aseptic work easier in routine water testing.
- Method suitability: Review the filter conditions specified by your laboratory method.
Frequently asked questions
Why is a 0.45 µm pore size used most often?
Most of the bacteria targeted in water testing are retained on the filter surface at this pore size, while the filtration speed stays practical. Still, the correct pore size is the value stated by the method you apply.
What should I watch for when placing the filter on the medium?
The gridded side must face up and no air bubbles should remain between the filter and the agar. Where a bubble remains, the medium cannot reach the filter and colonies may not grow; handle the filter only with flat-tipped forceps.
How do I order Diaplas Membrane Filter 47 mm?
For bulk orders, size options and quotes, contact us on +90 216 473 80 70, via our WhatsApp line or through the quote form; our team will promptly send you the most suitable option and a quote.