What Is a Stomacher Bag and How Is It Used? Selection Guide

A stomacher bag (homogeniser bag) is a single-use sample bag used in a paddle blender to homogenise food, feed, surface or water samples in a diluent inside a closed, sterile space. Because the sample stays inside the bag and never touches the paddles of the instrument, the risk of cross contamination between samples is largely removed and no parts have to be washed after each analysis.

What is a stomacher bag used for?

In microbiological analysis the microorganisms in a sample have to pass into the liquid phase. This is achieved by homogenising the sample together with a diluent, which yields the initial suspension that starts the series of decimal dilutions. The stomacher bag is the vessel for that homogenisation: it serves as weighing container, mixing chamber and pipetting surface at the same time.

Working in a closed system has two practical benefits. First, contamination of the laboratory environment through aerosols and splashes is reduced. Second, since the instrument does not need cleaning, the move to the next sample is faster. This is why stomacher bags are among the most basic consumables in food microbiology laboratories.

How does a paddle blender work?

A paddle blender holds the bag between a fixed plate and two moving paddles. The paddles compress the bag in turn; each stroke crushes the sample, disperses it in the diluent and carries cells into the liquid phase. Unlike blade-type mixers there is no rotating blade, so the sample does not overheat, cell damage stays more limited and the instrument does not need to be sterilised.

Run time depends on the structure of the sample. Soft, watery samples disperse quickly, while fibrous or fatty samples need somewhat longer; for most foods a period of roughly 30 seconds to 2 minutes is sufficient. Homogenising for longer than necessary can warm the sample and reduce the viability of some sensitive microorganisms.

Difference between filter and non-filter stomacher bags

CriterionNon-filter bagFilter bag
StructureOne single chamberTwo compartments separated by a non-woven filter
Liquid that is pipettedMay contain particlesClear liquid freed from coarse particles
Blocking of pipette tipsCommon with fibrous and granular samplesMarkedly reduced
Suitable samplesLiquids, purees and easily dispersed productsDifficult samples such as meat, fatty products, flour, spices, fibrous vegetables and cheese
Molecular and automated systemsAn extra clarification step may be neededMore suitable for PCR and automated pipetting
CostMore economicalHigher because of the filter

The selection rule fits in one sentence: if the clarity of the liquid you pipette affects the result, use a filter stomacher bag; if it does not, a classic stomacher bag is enough. Since filter porosity differs between products, the product information should be checked for applications in which very fine particles are expected to be retained.

How is the initial dilution prepared according to ISO 6887-1?

  1. The diluent is prepared and brought to working temperature; as a general rule it should be at room temperature.
  2. The bag is placed in a sterile rack or weighing frame, and its mouth is folded outwards so that the outer surface is not soiled.
  3. The sample is weighed aseptically and diluent is added to give a ratio of 1 in 10; for example 90 ml for 10 g of sample, or 225 ml for 25 g.
  4. Air is expelled from the bag, the bag is closed or folded and clamped, and it is placed in the paddle blender.
  5. The instrument is run for a time appropriate to the sample type. At the end no visible large pieces should remain in the bag.
  6. The decimal dilution series is prepared by pipetting from the suspension in a non-filter bag, or from the clear side of the filter in a filter bag.
  7. The interval between preparing the initial suspension and inoculation is kept as short as possible; waiting can change the results.

Which diluent should be chosen?

  • Buffered peptone water (BPW): used both as initial suspension and as pre-enrichment broth for samples in which Salmonella is sought. The ready-to-use BPW in bottles format is practical for large sample volumes.
  • Maximum Recovery Diluent (MRD): a general purpose diluent containing peptone and sodium chloride that protects stressed cells; it is widely used in enumeration work. The ready-to-use MRD in bottles option suits the initial suspension.
  • Saline peptone water: another general diluent used for a similar purpose, available in bottles and in tubes.
  • Liquids containing neutralisers: where disinfectant residues are expected on surface and equipment samples, the residue has to be neutralised.

Which diluent is used is determined by the microorganism sought and by the standard applied. For the sampling stage as a whole, our guide to food sampling gives direction.

Bag volume, filling and closing

Bag volume is chosen according to the total volume of sample plus diluent and to the free movement of the paddles. A common practice is to work with bags classified as 400 ml for 25 g of sample and 225 ml of diluent; the bag is not filled completely, and headspace is left so that the paddles can move freely over the liquid.

The mouth of the bag can be closed by folding, with a clip or with the bag closure itself. If the bag escapes from the clamp during homogenisation, liquid splashes into the instrument and the sample is lost, so the closure should always be checked.

Common mistakes

  • Sharp and bony samples: a bone fragment, fish bone or hard shell can puncture the bag. For such samples a more resistant bag or placing the bag inside a second one should be considered.
  • Overfilling: the paddles cannot move the whole volume, homogenisation stays incomplete and the risk of overflow at the mouth increases.
  • Not expelling the air: air left in the bag acts as a cushion and prevents the sample from being crushed.
  • Running longer than necessary: warming and foam formation can lower recovery of sensitive microorganisms.
  • Misusing the filter: in a filter bag the sample goes into the compartment without the filter and pipetting is done on the clear side; doing the opposite cancels the function of the filter.
  • Unlabelled bags: sample identity is lost. Labelling before weighing, or writing on the marking area, should become a habit.

For a bag and diluent combination suited to your sample types, you can contact us.

Sources and standards

  • ISO 6887-1:2017 — Microbiology of the food chain — Preparation of test samples, initial suspension and decimal dilutions for microbiological examination — Part 1: General rules for the preparation of the initial suspension and decimal dilutions
  • ISO 6887-4:2017 — Microbiology of the food chain — Preparation of test samples, initial suspension and decimal dilutions for microbiological examination — Part 4: Specific rules for the preparation of miscellaneous products
  • ISO 7218 — Microbiology of the food chain — General requirements and guidance for microbiological examinations
  • ISO 6579-1:2017 — Microbiology of the food chain — Horizontal method for the detection, enumeration and serotyping of Salmonella — Part 1: Detection of Salmonella spp.

Questions

Frequently asked questions

What is a stomacher bag used for?

A stomacher bag allows a sample to be homogenised in a diluent inside a closed, sterile space during microbiological analysis. The microorganisms in the sample pass into the liquid phase and the initial suspension that starts the decimal dilution series is obtained. The bag also serves as weighing container and pipetting surface, and because the sample never touches the instrument the risk of cross contamination between samples is greatly reduced.

When should a filter stomacher bag be used?

A filter bag is used when the liquid being pipetted has to be clear. With samples such as meat, fatty products, flour, spices, fibrous vegetables and cheese, liquid taken from a non-filter bag contains particles; these block pipette tips and make molecular methods or automated pipetting difficult. The filter holds coarse particles in one compartment and allows pipetting from the clear side.

How much sample and diluent go into a stomacher bag?

The initial suspension is usually prepared at a ratio of 1 in 10: 90 ml of diluent for 10 g of sample, or 225 ml for 25 g. Bag volume is chosen so that this total volume fits comfortably, with headspace left for free paddle movement; the bag is never filled to the mouth. Where a sample type requires a different ratio, the relevant part of the applied standard governs.

Which diluent is used in a stomacher bag?

The choice depends on the microorganism sought and the standard applied. For samples in which Salmonella is sought, buffered peptone water (BPW) serves both as initial suspension and as pre-enrichment broth. In general enumeration work Maximum Recovery Diluent (MRD) or saline peptone water is common. For surface samples where disinfectant residues are expected, a liquid containing a neutraliser is preferred.

How can stomacher bags and diluents be obtained from Diaplas?

For filter and non-filter stomacher bags and for ready-to-use diluents such as BPW, MRD and saline peptone water you can reach us on +90 216 473 80 70 or fill in the quotation form on the website. If you share your sample types, your daily sample numbers and the homogeniser model you use, we can determine a suitable bag volume and diluent format together.